Case Report
Cardiac Tamponade Caused by Fracture and Migration of Inferior Vena Cava Filter
A 53-year-old male presented to the emergency room with acute cardiopulmonary compromise. Echocardiography revealed cardiac tamponade. The patient was taken emergently for surgery and a 28 gauge wire of 1.5 cm was retrieved from his right ventricle. A section had fractured from the inferior vena cava (IVC) filter and migrated…