Primary Article

Detecting Ductal Shunting in Premature Infants by Range-Gated Doppler Echocardiography

Authors: ALLAN REES MD, FRANCISCO ELBL MD, ROBERT SOLINGER MD, JOHN GATTINELLA CPT, LYDIA LAMBERT LPN, LARRY COOK MD

Abstract

Range-gated pulsed Doppler echocardiography has been reported to be a useful noninvasive bedside technic for detecting ductal left-to-right shunting in premature infants. We studied 30 premature infants with a birth weight of less than 1,500 gm, using a 5 MHz system developed by Advanced Technology Laboratories. An umbilical artery catheter was used to obtain a contrast aortogram. Twenty-three infants had both a positive flush aortogram and a continuous turbulence documented by Doppler echocardiography. A clinically audible murmur was present in 17 of these patients. In six infants there was no Doppler evidence of PDA. In five of this last group there was also a negative flush aortogram. The excellent correlation found between the Doppler and aortographic studies suggests that this noninvasive technic is very sensitive for detecting ductal left-to-right shunting, even in the absence of an audible murmur.

This content is limited to qualifying members.

Existing members, please login first

If you have an existing account please login now to access this article or view purchase options.

Purchase only this article ($25)

Create a free account, then purchase this article to download or access it online for 24 hours.

Purchase an SMJ online subscription ($75)

Create a free account, then purchase a subscription to get complete access to all articles for a full year.

Purchase a membership plan (fees vary)

Premium members can access all articles plus recieve many more benefits. View all membership plans and benefit packages.

References